Chapter 8 - Replacement Procedures
8-95
8) Working from the
left
side of the unit, release the swivel release cable [
A
] from the cable retaining
bracket [
B
] located on the vertical axis welding [
C
] by pressing the tensioning plate [
D
]
downwards
(to release tension on the cable). Then slide the cable end stopper [
E
] and cable sleeve [
F
] out of
the slot in the retaining bracket.
9) Working from the
front
of the unit, and using an M4 Allen key, loosen the two cap screws on the
swivel release cable retaining bracket and slide the bracket
forwards
sufficiently to allow removal
of the cable from the up/down/swivel handle, as shown in
Figure 8-77 Swivel Release Cable and Vertical Axis Welding (from Side)
Figure 8-78 Releasing the Cable End Stopper from the Tensioning Plate
C
D
B
A
E
F
F
D
B
A
E