Chapter 3 - Installation
3-47
3-6-10
Technical Support Tab
The System Configuration
Technical Support
tab is used when new hardware has been installed, or
when it necessary to enable new functionality. The appearance of the Technical Support tab differs in
accordance with the specific Vivid™ 4 BT03 model (RFI or RFT configuration), as described below.
3-6-10-1
Technical Support Tab - RFI Configuration
Monitoring
- should
always
be enabled (this provides constant monitoring of critical
parameters to ensure safe operation of the system).
Watchdog
- normally should not be enabled. When enabled, if there is a system lock-up
caused by the Back End, this will turn OFF power at the Front End.
MOD Fan
- should be enabled only on BEPs based on Pentium 4 - this provides faster
operation of the MOD. Enabling this option in the wrong type of BEP may
cause damage to the MOD or MO media.
Use Front End
- normally should be enabled.
ILinq
- should be enabled only on systems that use the iLinq option.
Figure 3-29 Technical Support Tab - RFI Configuration