Chapter 3 - Installation
3-65
5) Press
<Ctrl+S>
to view the status of the spooler, as shown below:
Figure 3-47 Job Spooler
The
Status
column can contain the following options:
•
Hold
:
Images are in the spooler buffer but transmission has not begun as the exam is
active.
•
Active
:
Images are being transmitted over the network.
•
Done
:
Transmission is complete and all the images have been transmitted over the
network.
•
Fail
:
Transmission failed or is incomplete. An error message will be displayed in the
Error
column.
3-7-6
Connecting Directly to EchoPAC (or Image Vault)
Note:
The EchoPAC station will reject any images that have a future date. Ensure that the clock on
the Vivid™ 4 ultrasound unit lags behind that of the EchoPAC system by a minute or two.
1) Press
Config
on the alphanumeric keyboard. The
System Configuration
window is displayed.
2) Trackball to the
Connectivity
tab and press
Select
. The
Connectivity
tab is displayed, with the
Views
subsidiary tab selected by default, as shown in
3) Trackball to the
Dataflow
button and press
Select
. The full list of data flows is displayed, as shown
in
4) Expand the
Remote Archive - Remote Storage
branch. A service called
Remote Servers
is
displayed.
5) Expand the
Remote Servers
branch. A service called
Remote Archive - Remote Storage
is
displayed.
6) Trackball to the
Services
tab and press
Select
. The
Services
tab is displayed, as shown in