GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
10-10
Section 10-3 - Periodic Maintenance Schedule
10-5-4
Cleaning
10-5-4-1
General Cleaning
10-5-4-2
Air Filter Cleaning
Regular cleaning of the Vivid™ 4 scanner’s air filter will ensure that the filter does not become clogged
- this would otherwise cause the system to overheat and reduce system performance and reliability.
NOTE:
It is recommended that the air filter cleaning procedure is performed once every 3 months (quarterly).
However, the required frequency for air filter cleaning will vary in accordance with environmental
conditions.
Table 10-8 General Cleaning
Step
Item
Description
1
External
Panels
On a weekly basis, moisten a soft, non-abrasive folded cloth with a mild, general purpose,
non-abrasive soap and water solution or general purpose disinfectant. Wipe down the top,
front, back and both sides of the unit
.
Do not spray any liquid directly onto the unit!
2
Monitor
On a weekly basis, apply glass cleaner to a soft, non-abrasive folded cloth. Gently wipe the
monitor face.
3
Console
Frequent and diligent cleaning of the Vivid™ 4 ultrasound unit reduces the risk of spreading
infection from person to person, and also helps to maintain a clean working environment.
Prior to cleaning, turn OFF the power to the system.
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system.
Be careful not to get the cloth too wet so that moisture does not enter the console.
4
Control Panel
and Keyboard
Control Panel:
On a weekly basis, moisten a soft, non-abrasive folded cloth or sponge with a mild, general
purpose, non-abrasive soap and water solution or general purpose disinfectant. (
Do not use
any solution containing abrasive powder or strong chemicals such as, acid or alkaline
).
Squeeze excess liquid from the cloth/sponge, them wipe down the Control Panel
.
Do not spray any liquid directly onto the Control Panel!
Rinse the cloth/sponge with clean running water and wipe the Control Panel again.
Use a dry, soft, lint-free cloth to dry the Control Panel.
Wait for the Control Panel surface to dry completely.
Keyboard:
Clean the keyboard as described (above) for cleaning the Control Panel.
Note: In the event that disinfection is required or any stubborn stains remain, absorb a small
quantity of isopropyl rubbing alcohol on a soft, dust-free cloth.
Wipe the surface of the keycaps with the cloth, making sure that no liquid drips on or between
the keys. Allow to dry.
5
Probe Holder
Clean the probe holders with warm water and a damp cloth to remove all traces of gel
(soaking may be required to remove excess gel).