GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
Chapter 10 - Periodic Maintenance
10-23
10-7-5-3
Data Sheet for Chassis Source Current Leakage
The Chassis Source Current Leakage Test passes when all readings measure less than the values shown
in
Vivid™ 4 Preventative Maintenance Inspection Certificate
(see
Table 10-17 Typical Data Sheet for Chassis Source Current Leakage Test
Unit
Power
Tester
Polarity
Switch
Tester
Neutral or
Ground
Switch
Test 1
Probe
Connector
Test 2
Caster
Wheel
Test 3
CRT
Test 4
(Optional)
Test 5
(Optional)
Enter Name of tested peripheral here:
ON
NORM
OPEN
ON
NORM
CLOSED
ON
REV
OPEN
ON
REV
CLOSED
OFF
NORM
OPEN
OFF
NORM
CLOSED
OFF
REV
OPEN
OFF
REV
CLOSED