GE M
EDICAL
S
YSTEMS
D
IRECTION
2392751-100, R
EVISION
3
V
IVID
™ 4 S
ERVICE
M
ANUAL
8-78
Section 8-5 - Back End Parts Replacement
17.)Disconnect the power cable from the Floppy Disk (labelled [
6
); refer also to
18.)Disconnect the mains power cable (4-pin flat connector [
7
] - Part No. 2252981) from the rear of the
BEP, cutting any cable ties as necessary.
19.)Disconnect the mains power cable [
8
] from the motherboard (beside the ECG location).
20.)Gently slide the power supply further downwards and then out of the BEP casing sufficiently to be
able to cut any remaining cable ties.
21.)Remove the power supply from the BEP, together with the attached cable loom, as shown in
22.)Disconnect the 4-pin external connector [
9
] from the motherboard (beside the CPU).
Figure 8-63 Removing the BEP2 Power Supply with Cable Loom and Connectors