BreathSpec®
– User Manual
87/242
Factory default limits
Name
Lower limit
(%)
Upper limit
(%)
EPC2-Pressure (kPa)
95
110
RIP Height (V)
80
150
RIP Position at 101.33 kPa (ms)
95
105
FWHM (ms)
80
120
INFORMATION!
The device is delivered with an acceptance snapshot. This
snapshot defines the system performance during device
acceptance and is used to assess the readiness for measurement.
Any deviations from this are displayed in the Error Information
Window.
The default values can be adjusted by the customer
.