BreathSpec®
– User Manual
120/242
6
After all temperatures
reached the default
values inspect visually
the current spectrum and
compare it with the
reference spectrum of the
analytical approval.
The
RIP should reach ~80%
of the RIP height
displayed in the
delivered Analytical
Approval of the device
under same
measurement
conditions of G.A.S.
The assessment of the
readiness to measure is
additionally supported
by the system.
If not check the gas
quality and/or install
additional purification
cartridges and start the
cleaning procedure
again.
INFORMATION!
The device is delivered with an acceptance snapshot. This
snapshot defines the system performance during device
acceptance and is used to assess the readiness for measurement.
Any deviations from this are displayed in the Error Information
Window.
The default values can be adjusted by the customer
.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...