BreathSpec®
– User Manual
174/242
9.16 Workflow: Start Standby mode
INFORMATION!
It is recommended not to switch off the device during measurement
breaks. Using the Standby mode ensures the cleanness of the system
and a quick readiness to measure.
1
Select
System
›
Modes
›
Standby [x h]
and set the
standby duration time.
Select
Start
to start the
Standby process.
Driftgas (EPC1) is set to
10 ml/min
and Carriergas
(EPC2) is set to
5 ml/min
.
2
During the standby
process a Standby
Window appears.
During this process no
other activities can be
executed.
This process can be
skipped at any time.