9.19 Workflow: Change sample pump unit
INFORMATION!
Only authorized, trained and technically instructed people are
allowed to do these work steps.
DANGER!
Before all work on the device
switch off the device
and
pull out the
power plug!
1
Switch off the device and
pull out the power plug.
2
Remove the top cover.
For detailed information
see
Workflow: Remove
Housing Cover
.