BreathSpec®
– User Manual
30/242
Figure 2: Ion Mobility Spectrometer -Working Principle
Atmospheric Ionization of molecules can be obtained by several techniques. G.A.S.
uses photoionization with a 10.6eV UV-lamp or soft chemical-ionization initiated by a
low-radiation tritium (H3) source (below exemption limits of EURATOM). While the
first directly produces positive ions, the latter generates
reactant ions
with the gas
atmosphere by a cascade of reactions following the collision of a fast electron
emitted from the
β
-radiator H3. The so-called Reaction Ion Peak (RIP) representing
the total of all ions available is formed as a first step. In nitrogen and air, resp., the
reactant ions
can be described as H+(H2O)
n
and O2-(H2O)
n
. Chemical ionization of
analytes by
reactant ions
then result in the formation of specific analyte ions, when
the affinity of the analyte towards the reactant ion is higher in case compared to
water (using the positive ionization mode). The proton affinity of water is 691kJ/mol,
so all molecules with a higher proton affinity will be ionized by a proton transfer,
which is typically given for all heteroatom-organic compounds. The ionization takes
place at ambient pressure, so that the analyte concentration is not diluted as
compared to other analytical methods where a vacuum must be applied. Therefore,
IMS is extremely sensitive. The detection limits typically are in the low ppb-range for
volatile organic compounds
(VOC).
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...