BreathSpec®
– User Manual
195/242
8
The controller finds a
position again and the
switching noise is
normal
again.
INFORMATION!
If the valve cannot find a position or the switching noise does not return
to normal, repeat the procedure.
9.22 Workflow: Change filter set CGFU
INFORMATION!
It is recommended change the filter set of the Circular gas flow unit
periodically. The recommended maintenance interval is every
6
month
.
DANGER!
Before all work on the device switch off the device and pull out the
power plug!
1
Switch off the device and
pull out the power plug.