BreathSpec®
– User Manual
230/242
Symtom
Find signals from the prior measurement run in the
chromatogram
Possible Cause
The measurement runtime is too short.
Action
Increase the measurement run time
Symtom
The actual size of the measurement file is to large
Possible Cause
Average setting too low.
Action
Increase the Average setting or shorten the runtime
(recommended averaging: 6).
Symtom
The measurement signals cannot mapped well.
Possible Cause
Average setting is too high.
Action
Decrease the Average setting (recommended averaging: 6).
Symtom
EPC2 Maximum flow of 150 ml/min is not reached.
Possible Cause
The reachable maximum flow depends on the dimensions of
the installed column.
Action
Get the specified maximum flow from Analytical approval. If
that still does not help contact the G.A.S. service hotline.
Possible Cause
When using a CGFU unit the total flow (EPC1 and EPC2) is
limited to 400ml/min.
Action
No Action
Symtom
Six-Port-
valve doesn’t switch
Possible Cause
Lost connection
Action
Restart the system. If that still does not help contact the
G.A.S. service hotline.
Symtom
Six-Port-valve switching noise sounds strange
Possible Cause
Initialization lost
Action
Reintializise the Six-Port-Valve
Symtom
No sample signals are displayed during a measurement.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...