BreathSpec®
– User Manual
225/242
Action
The device get a wrong trigger signal. Check the external
trigger programming.
Repeat the procedure.
If that still does not help contact the
G.A.S. service hotline.
Error message
Stop recording first.
Description
When starting the trigger mode recording is activated.
Action
Deactivate recording and repeat the procedure.
If that still
does not help contact the G.A.S. service hotline.
Error message
Select program first.
Description
When starting the trigger mode no program is activated.
Action
Select a program and repeat the procedure.
If that still does
not help contact the G.A.S. service hotline.
Error message
Can’t read calibration.gsd
Description
The import of the calibration.gsd. file failed. The file was not
found.
Action
Ensure that the file calibration.gsd exist and is placed in the
measurement subfolder of the device. Check network cable,
network shares and IP-address.
Repeat the procedure.
If that still does not help contact the
G.A.S. service hotline.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...