BreathSpec®
– User Manual
32/242
Complex analyte mixtures, like e.g. food flavours, often demand a second and
independent separation step to separately analyse the multiplicity of compounds at
low concentrations. Therefore G.A.S. -according to application- equips its IMS
systems with gas chromatographic (GC) columns. The volatile compounds of
samples under testing are pre-separated in time by a GC column. The discrete
compounds are consecutively fed into the IMS ionization chamber, so that analyte
and/or ion interactions are avoided.
Furthermore, a competition of analytes on the reactant ions is excluded, enhancing
the sensitivity of the system for individual compounds.
The GC-IMS setup enables a twofold separation of analyte mixtures and the
detection by the IMS electrometer. Since the IMS measurements are extremely fast
(30ms / spectrum) a continuous and high-resolution recording of analyte signals is
provided.
Figure 5: GC-IMS measurement 3D
The above figure sketches the GC-
IMS measurement‘s 3D-dataset and the
corresponding heatmap visualization.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...