9.3.4 Workflow: Execution of an ambient air measurement using the Handheld
The analysis of the ambient air may indicate volatile compounds that do not originate
from the subject/patients breath but are present in the air.
To analyze the ambient air follow the steps described in 9.3.2. A Reservoir tube and
Mouthpiece must not be installed.
9.4 Workflow: Breath Sampling via Disposable Breath Sampler
9.4.1 Start a breath measurement using disposable breath sampler
INFORMATION!
To sample via the Luer-Lock-Adapter the respective tube port has to be
installed
see chapter 7.5
.
DANGER
•
Biohazard: wear appropriate gloves at all handling steps
•
Take measures to prevent contamination from one subject/patient
to other
WARNING
•
Pinching possible
1
Press Start Button to
initialize
the
selected
program.
The
dialog
‘Start progam’
appears.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...