BreathSpec®
– User Manual
223/242
one. Repeat the procedure.
If that still does not help contact
the G.A.S. service hotline.
Error message
No measurement files transferred
Description
No measurement are transferred to an external storage.
Action
Check the external storage. Check network cable, network
shares and IP-address. Check write protection of the USB-
Stick. Ensure that the USB-Stick is FAT32 formatted.
Reformat the USB-Stick. Replace the USB-Stick with a new
one. Repeat the procedure.
If that still does not help contact
the G.A.S. service hotline.
Error message
No measurements stored.
Description
No measurements were deleted, because no measurements
are available.
Action
Do a measurement. Repeat the procedure.
If that still does
not help contact the G.A.S. service hotline.
Error message
RTC read error.
Description
Date and time setting failed.
Action
Restart the system and repeat the procedure.
If that still does
not help contact the G.A.S. service hotline.
Error message
Save settings failed.
Description
Setting data could not be save on a USB-stick.
Action
Check write protection of the USB-Stick. Ensure that the
USB-Stick is FAT32 formatted. Reformat the USB-Stick.
Replace the USB-Stick with a new one. Repeat the
procedure.
If that still does not help contact the G.A.S.
service hotline.
Error message
Load settings failed.
Description
Setting data could not be load from a USB-stick
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...