BreathSpec®
– User Manual
29/242
5 Introduction
5.1 Working principle of IMS technologie
Ion Mobility Spectrometry (IMS) is an analytical technology to separately detect
gaseous compounds in a mixture of analytes. The separation is based on the specific
drift times, that ionized compounds need to pass a fixed distance (drift tube) in a
defined electric field.
Figure 1: Ion Mobility Spectrometer - Basic Relations
Compared to other techniques e.g. TOF-MS, ions travel at atmospheric pressure
versus a flow of inert drift gas. The drift time of each substance is determined by its
ion mass and geometric structure, as slowing collisions with the drift gas molecules
are more frequent for sterically demanding structures. Therefore, IMS can even
differentiate isomeric molecules. For detection, the resulting ion current is measured
by an electrometer as a function of time.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...