6.7.3.1 Snapshot Window
The instrument performance and status can be monitored using recorded snapshots.
It is an automatic comparison of current instrument settings versus the operator
defined target settings which helps the operator to assess the readiness of the
system for starting a measurement. This automatic monitoring can recognize among
other a System contamination, insufficient gas quality or system leaking. For this
purpose, the following parameters are constantly cross-checked against the target-
settings:
•
The carriergas pressure - EPC2 pressure (kPa)
•
The height of the Reaction-Ion-Peak (RIP) - Rip-height (V)
•
The position of the Reaction-Ion-Peak (RIP) normalized to normal pressure
–
Rip Pos at 101.33 kPa (ms)
•
The full width at half maximum of the Reaction-Ion-Peak - FWHM (ms)
•
The temperature values T1-T6
•
The gasflow and pressure of Driftgas (EPC1) and Carriergas (EPC2)
The user can change the default factory settings for his needs. Error messages are
displayed in the Error Information Window.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...