Error codes
Bit
Value
Information
0
No Error
0
1
Unknown Error
. Source of
Error unknown.
1
2
Error concerning
current
.
Only used in special
applications.
2
4
The
temperature
of a
component has not reached
the designated value
3
8
The
flow
of one or more
pressure controllers has not
reached the designated value
4
16
Storage
problems. E.g.
cannot write to the device
15
32768
Multiple Trigger
Status codes
Bit
Value
Information
0
1
Concentration
within range
1
2
Concentration is
below
the
calibrated range
2
4
Concentration is
above
the
calibrated range
3
8
Invalid
3-15
16-32768
Unused
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...