Error message
Unable to unmount USB-Storage.
Description
Unmounting of the USB-Stick failed.
Action
Remove the USB-Stick and restart the device. Check the
USB-Stick, optionally reformat it. Insert the USB-Stick and
repeat the procedure. If that still does not help contact the
G.A.S. service hotline.
Error message
No program selected.
Description
At program start no program was detected
Action
Create and select a program. Repeat the procedure.
If that
still does not help contact the G.A.S. service hotline.
Error message
Drift gas flow too low. Aborting…
Description
At program start the drift gas flow is too low. Programstart
will be refused.
Action
Increase the back pressure. Repeat the procedure.
If that still
does not help contact the G.A.S. service hotline.
Error message
Stop recording first.
Description
If recording is activ a programstart is refused.
Action
Deactivate recording and repeat the procedure.
If that still
does not help contact the G.A.S. service hotline.
Error message
Can’t initalize drift supply. Program can not be started.
Description
Drift voltage error
Action
Fatal Error! Contact the G.A.S. service hotline.
Error message
Invalid program.
Description
The selected program has no instructions.
Action
Complete the program and repeat the procedure.
If that still
does not help contact the G.A.S. service hotline.
Error message
Not enough storage space left. Please delete or copy
measurement or choose a shorter program.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...