BreathSpec®
– User Manual
81/242
Snapshot
The
Manage
Button opens
the
snapshot
window. User-specific target values can
be set. A new snapshot can be recorded
or can be imported
Snapshot is a new function to allow
automatic instrument performance
checking
Contamin
Define the automatic search area for
contamination. The search window can
be defined by setting the Threshold
(signal level in mV which is recognised as
a contamination) and the Start of the
search in the drift-time spectrum given as
RIP relative proportional factor.
•
Current:
Shows the currently
detected contamination Shows the
currently detected contamination.
•
Threshold:
The set value from
which an error is triggered.
Search from:
The RIP-relative starting
position of the search.
Current Loops
(optional)
Activating the
Settings Button
opens a
current loop setup dialog (only visible if
available)
Live Spectra:
Prepared for further use! Currently not in
function!
Live Sensor:
The settings for Live Data communication
with G.A.S. Sequence Designer Software
can be set. For detailed Information refer
to the G.A.S. Sequence Designer
Software Manual.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...