BreathSpec®
– User Manual
122/242
INFORMATION!
To ensure correct measurements it is absolutely necessary to
connect the supplied exhaust tubes
(Gas out and Sample gas
out).
The exhaust tubes (Gas out and Sample gas out)
must be led
separately into the exhaust system and must not be
connected.
The exhaust system must
not generate any negative pressure
.
8.2 Installing the device
1
Unpack the device.
Remove the foam
spacers.
Remove the accessories
boxes.
Lift the device from the
transport box.
WARNING!
Do not deploy the device in areas exposed to explosive substances or
mixtures!
2
Place it on a robust table
top of minimum footprint
size of 60 cm x 70 cm.
Ensure that there is
enough working space at
the rear side of the
device.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...