SS1807-N002
Installation
-
39/68
Function
Description
FICE
This switch changes the ON/OFF of FICE.
When a FICE set number is assigned in the “Simple FICE Setup” and
FICE is “OFF”, each pressing of this switch changes the function to be
selected in order of FICE “ON” and call up of FICE set number of
“Assignment 1” → call up of “Assignment 2” → call up of “Assignment 3”
→ FICE “OFF”.
The “Simple FICE Setup” is set by service personnel.
→EP-6000 Operation Manual “5.3.3 Setting FICE”
Color Emphasis
This switch turns the tone function ON/OFF.
→EP-6000 Operation Manual “5.3.4 Tone Settings”
Display
This switch is used to display or not to display the information on the
observation screen.
* When “Data Display Switching” is disabled, the function of this switch is
also disabled.
Electronic Zoom
When this switch is pressed, the display size changes to the size specified
in “Zoom Ratio of Electronic Zoom”. When it is pressed again, the display
size returns to the original size (zoom ratio: x1.0).
* The “Zoom Ratio of Electric Zoom” is set by service personnel.
Optical Zoom:
Zoom In
When an optical zoom scope is connected, this switch changes the zoom
ratio of the optical zoom. This switch performs zoom-in only.
Optical Zoom:
Zoom Out
When an optical zoom scope is connected, this switch changes the zoom
ratio of the optical zoom. This switch performs zoom-out only.
SU Freeze
When this switch is pressed, a freeze request is sent to the ultrasonic
processor. This function is available only when “SU-1” is selected in the
peripheral device setup.
* This function cannot be assigned to the scope switches on endoscopes
other than ultrasonic endoscopes.
SU Store
When this switch is pressed, a store request is sent to the ultrasonic
processor. This function is available only when “SU-1” is selected in the
peripheral device setup.
* This function cannot be assigned to the scope switches on endoscopes
other than ultrasonic endoscopes.
Timer
This switch starts or stops the timer.
→EP-6000 Operation Manual “7.17.2 Starting/Stopping the Timer”
Lap Time
When this switch is pressed, the lap time displayed on the observation
screen is started/stopped/reset.
→EP-6000 Operation Manual “7.17.3 Lap Time Function”
Not Assigned
Assigns no function.
Содержание EP-6000
Страница 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Страница 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Страница 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Страница 18: ...SS1807 N002 Caution in Safety 6 12 2 Label 2 1 EP 6000 Labeling Layout FV693A 2 1 1 Labeling chart D B A C ...
Страница 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Страница 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Страница 52: ...SS1807 N002 Instruction of System 17 106 In the case of Normal mode In the case of BLI BLI bright or LCI ...
Страница 131: ...SS1807 N002 Instruction of System 96 106 9 EP 6000 Description of Configuration 9 1 Block Diagram ...
Страница 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Страница 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Страница 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Страница 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Страница 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Страница 348: ...SS1807 N002 Service Parts List 6 19 6 13 13 16 18 17 17 6 18 14 15 6 9 10 6 12 7 6 11 8 6 5 1 4 3 2 ...
Страница 350: ...SS1807 N002 Service Parts List 8 19 13 12 14 11 14 17 7 7 2 6 15 7 2 3 4 1 5 2 10 9 17 7 7 8 18 12 13 16 ...
Страница 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Страница 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Страница 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Страница 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Страница 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Страница 377: ...SS1807 N002 Installation 1 68 Installation ...
Страница 381: ...SS1807 N002 Installation 5 68 1 2 Installation onto the Cart Standard System Installation Example ...
Страница 445: ...SS1807 N002 ...