SS1807-N002
Failure Analysi
-
14/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
130
Information
In the process of
acquiring scope operating
mode.
Please wait for a while.
The operating mode selection
screen in the factory menu was
opened.
Same as on the left.
The message was displayed
when trying to open the screen.
In case when acquisition of the
operation mode frequently fails
even after waiting for a while,
check the version of the Rear
CPU and Front CPU. If the
version is later than 580 60P
support, replace the patient PCB.
131
Information
In the process of setting
scope operating mode.
Please wait for a while.
The setup button was pressed on
the operating mode selection
screen in the factory menu.
Same as on the left.
The message was displayed
when trying to set the operating
mode.
In case when acquisition of the
operation mode frequently fails
even after waiting for a while,
check the version of the Rear
CPU and Front CPU. If the
version is later than 580 60P
support, replace the patient PCB.
132
Information
Please enter password for
forwarding correction
function.
133
Information
Please enter current
password.
(Press [ESC] key to
cancel. )
134
Information
Please enter a new
password (four or more
characters).
(Press [ESC] key to
cancel. )
135
Information
Please enter a new
password once again for
confirmation.
(Press [ESC] key to
cancel. )
136
Information
Please enter a password.
(Press [ESC] key to
cancel. )
137
Information
Please enter a password.
(Press [ESC] key to
cancel. )
138
Information
Different password.
Please enter the correct
password.
139
Information
Different password.
Please enter the correct
password.
Содержание EP-6000
Страница 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Страница 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Страница 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Страница 18: ...SS1807 N002 Caution in Safety 6 12 2 Label 2 1 EP 6000 Labeling Layout FV693A 2 1 1 Labeling chart D B A C ...
Страница 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Страница 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Страница 52: ...SS1807 N002 Instruction of System 17 106 In the case of Normal mode In the case of BLI BLI bright or LCI ...
Страница 131: ...SS1807 N002 Instruction of System 96 106 9 EP 6000 Description of Configuration 9 1 Block Diagram ...
Страница 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Страница 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Страница 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Страница 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Страница 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Страница 348: ...SS1807 N002 Service Parts List 6 19 6 13 13 16 18 17 17 6 18 14 15 6 9 10 6 12 7 6 11 8 6 5 1 4 3 2 ...
Страница 350: ...SS1807 N002 Service Parts List 8 19 13 12 14 11 14 17 7 7 2 6 15 7 2 3 4 1 5 2 10 9 17 7 7 8 18 12 13 16 ...
Страница 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Страница 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Страница 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Страница 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Страница 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Страница 377: ...SS1807 N002 Installation 1 68 Installation ...
Страница 381: ...SS1807 N002 Installation 5 68 1 2 Installation onto the Cart Standard System Installation Example ...
Страница 445: ...SS1807 N002 ...