SS1807-N002
Failure Analysi
-
19/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
213
Warning
Internal Memory has a
problem and cannot be
used.
Initialize Internal Memory.
Cause: Internal
Memory is not
initialized or its file
system is
corrupted.
When access was tried with the
setting of using an internal
memory, the states of internal
memory are as follows.
(1) Unformatted
(2) The file system was broken
Same as on the left
After performing backup of the
contents of internal memory in
Windows PC, external memory,
etc., initialize the internal memory
and then try back up again.
In case of repetitive errors after
that, replace the main PCB.
214
Warning
Internal Memory is full.
Cause: Free
space of Internal
Memory is
insufficient.
When access was tried with the
setting of using an internal
memory, data cannot be saved
due to insufficient internal
memory capacity.
Saving in internal memory was
tried.
After performing backup of the
contents of internal memory in
Windows PC, external memory,
etc., initialize the internal memory
and then try access again.
In case of repetitive errors after
that, replace the main PCB.
216
Warning
Some data is not yet
saved.
Save these data in
Internal Memory?
Some data is not yet saved.
・
Some data is not yet saved at
the time when the inspection
finished.
・
Some data is not yet saved at
the time when external memory
was inserted.
・
Some data is not yet saved at
the time when external memory
was initialized.
If displayed even if there is no
unsaved data, replace the main
PCB.
218
Warning
Saving print screen image
failed.
Check External Memory.
Saving in external memory has
failed.
When executing print screen
After backing up the contents of
inserted external memory in
Windows PC, external memory,
etc., perform physical-formatting
instead of quick-formatting of
external memory, insert the
external memory and try saving
again.
In case of repetitive errors after
that, replace the external
memory, USB PCB, main PCB.
219
Information
Free space of Internal
Memory is less than
100MB.
The free space of internal
memory became less than
101MB.
Same as on the left
In the case of repetitive errors
when the free space of internal
memory is 100MB or more, back
up the contents of internal
memory in Windows PC, external
memory, etc., initialize the
internal memory, and then check
again. In case of repetitive errors
after that, replace the main PCB.
Содержание EP-6000
Страница 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Страница 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Страница 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Страница 18: ...SS1807 N002 Caution in Safety 6 12 2 Label 2 1 EP 6000 Labeling Layout FV693A 2 1 1 Labeling chart D B A C ...
Страница 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Страница 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Страница 52: ...SS1807 N002 Instruction of System 17 106 In the case of Normal mode In the case of BLI BLI bright or LCI ...
Страница 131: ...SS1807 N002 Instruction of System 96 106 9 EP 6000 Description of Configuration 9 1 Block Diagram ...
Страница 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Страница 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Страница 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Страница 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Страница 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Страница 348: ...SS1807 N002 Service Parts List 6 19 6 13 13 16 18 17 17 6 18 14 15 6 9 10 6 12 7 6 11 8 6 5 1 4 3 2 ...
Страница 350: ...SS1807 N002 Service Parts List 8 19 13 12 14 11 14 17 7 7 2 6 15 7 2 3 4 1 5 2 10 9 17 7 7 8 18 12 13 16 ...
Страница 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Страница 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Страница 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Страница 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Страница 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Страница 377: ...SS1807 N002 Installation 1 68 Installation ...
Страница 381: ...SS1807 N002 Installation 5 68 1 2 Installation onto the Cart Standard System Installation Example ...
Страница 445: ...SS1807 N002 ...