Operating Instructions

Proline Promass E TB2

HART

Coriolis mass flow measuring system

BA01027D/06/EN/06.10
71112405

Valid as of version
01.00.zz (Device software)

Содержание Proline Promass E TB2 HART

Страница 1: ...Operating Instructions Proline Promass E TB2 HART Coriolis mass flow measuring system BA01027D 06 EN 06 10 71112405 Valid as of version 01 00 zz Device software...

Страница 2: ...s or the facility read the Basic safety instructions section carefully as well as all other safety instructions in the document that are specific to working procedures The manufacturer reserves the ri...

Страница 3: ...6 2 4 Turning the transmitter housing 22 6 2 5 Turning the display module 23 6 3 Post mounting check 24 7 Electrical connection 25 7 1 Preparing the measuring device for the electrical connection 25 7...

Страница 4: ...12 1 General troubleshooting 81 12 2 Diagnostics event on local display 82 12 2 1 Diagnostic message 82 12 2 2 Calling up remedial measures 83 12 3 Diagnostics event in the operating tool 84 12 4 Diag...

Страница 5: ...Version for experts Reference for your parameters The document provides a detailed explanation of each individual parameter in the 2nd part of the operating menu the Expert menu It contains all the d...

Страница 6: ...angerous situation Failure to avoid this situation can result in minor or medium injury NOTICE A0011192 EN NOTE This symbol contains information on procedures and other facts which do not result in pe...

Страница 7: ...ions that are forbidden A0011193 Tip Indicates additional information A0011194 Reference to documentation Refers to the corresponding device documentation A0011195 Reference to page Refers to the corr...

Страница 8: ...ed in the Operating Instructions and supplementary documentation the measuring device may be used for the following measurements only Measured process variables mass flow density temperature Calculate...

Страница 9: ...e not permitted and can lead to unforeseeable dangers If despite this modifications are required consult with Endress Hauser Repair To ensure continued operational safety and reliability Carry out rep...

Страница 10: ...nics module 7 Terminals pluggable spring terminals 8 Connection compartment cover 9 Sensor 3 2 Registered trademarks HART Registered trademark of the HART Communication Foundation Austin USA KALREZ VI...

Страница 11: ...Proline Promass E TB2 HART Product description Endress Hauser 11 3 3 Patents US patents EP patents 5 610 342 0 701 213 6 513 393 1 055 102 6 412 35 1 054 240 6 651 513 6 412 355 1 054 240 7 555 397...

Страница 12: ...e on the delivery note 1 identical to the order code on the product sticker 2 A0013695 A0013696 A0013698 Are the goods undamaged A0013696 i i DELIVERY NOTE A0013699 Do the nameplate data match the ord...

Страница 13: ...ww endress com deviceviewer 4 2 1 Nameplates Transmitter Order code Ext ord cd Ser no Date i i Patents 322540 0001 1 2 3 4 5 6 7 8 9 10 11 12 14 15 16 17 13 A0013906 2 Example of a transmitter namepla...

Страница 14: ...DG 3 A 14 Manufacturing date year month 15 Flow direction 16 2 D matrix code 17 Degree of protection e g IP NEMA 18 Approval information for explosion protection and Pressure Equipment Directive 19 Pe...

Страница 15: ...igher than the suspension points of the webbing slings Risk of injury if the measuring device slips Secure the measuring device from rotating or slipping Observe the weight specified on the packaging...

Страница 16: ...95 EC RoHS Packaging Wood crate treated in accordance with ISPM 15 standard which is confirmed by the affixed IPPC logo or Carton in accordance with European Packaging Directive 94 62EC recyclability...

Страница 17: ...ising from accumulation of gas bubbles in the measuring tube avoid the following mounting locations in the pipe Highest point of a pipeline Directly upstream of a free pipe outlet in a down pipe Insta...

Страница 18: ...iction mm in mm in 8 3 8 6 0 24 15 10 0 40 25 1 14 0 55 40 1 22 0 87 50 2 28 1 10 Orientation The direction of the arrow on the sensor nameplate helps to match the sensor with the direction of flow di...

Страница 19: ...precautions need to be taken for fittings which create turbulence such as valves elbows or T pieces as long as no cavitation occurs 19 Installation dimensions For the dimensions and installation leng...

Страница 20: ...he heat convection from the sensor to the transmitter A wide range of materials can be used to provide the required thermal insulation NOTICE Electronics overheating on account of thermal insulation O...

Страница 21: ...ess d 0 35 mm d 0 014 in Vibrations The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring system is not influenced by plant vibrations 6 1 3 Additio...

Страница 22: ...k 4 Remove stick on label on the electronics compartment cover 6 2 3 Mounting the measuring device WARNING Danger due to insufficient process sealing Ensure that the inside diameter of the seals is gr...

Страница 23: ...tment from the transmitter housing 3 Optional pull out the display module with a gentle rotation movement 4 Rotate the display module into the desired position Max 8 45 in each direction 5 Without dis...

Страница 24: ...perature range 19 Measuring range 97 m Has the correct orientation for the sensor been selected 18 According to sensor type According to fluid temperature According to fluid properties outgassing with...

Страница 25: ...able diameter Included cable glands M20 1 5 with cable 6 to 12 mm 0 24 to 0 47 in Plug in spring terminals for wire cross sections 0 5 to 2 5 mm2 20 to 14 AWG 7 1 3 Preparing the measuring device 1 Re...

Страница 26: ...0 mm 0 4 in For stranded cables also attach wire end ferrules 6 Firmly tighten the cable glands 7 2 1 4 3 A0013837 open 1 2 2 wire Flow Spare part 71108602 3 4 HART 4 20 mA 4 20 mA 04 05 HART 4 20 mA...

Страница 27: ...ments for the supply unit Supply voltage Order version output US supply voltage Imax output current 8E2B C 18 to 30 V DC 22 0 mA Supply voltage range without load For information on the supply voltage...

Страница 28: ...neatly and correctly Dry clean or replace the seals if necessary 2 Tighten all housing screws and screw covers 3 Firmly tighten the cable glands 4 So that any moisture that occurs does not reach the...

Страница 29: ...connection Endress Hauser 29 m If supply voltage is present is the device ready for operation and do values appear on the display module m Are all housing covers installed and firmly tightened m Is t...

Страница 30: ...ration options 8 1 Overview of operation options E 1 2 3 4 5 A0013902 1 Local operation via display module 2 Computer with operating tool e g FieldCare AMS Device Manager SIMATIC PDM 3 Field Xpert SFX...

Страница 31: ...king status Access status display Output n Operating menu for experts Language Display operat Access status display Locking status Operation Setup Wizard 1 Wizard 2 Wizard n Advanced setup Enter acces...

Страница 32: ...ntly pending diagnostic messages Event logbook submenu Contains up to 20 or 100 order option event messages that have occurred Device information submenu Contains information for identifying the devic...

Страница 33: ...ion check The device is in service mode e g during a simulation A0013958 Out of specification The device is operated Outside of its technical specifications e g during startup or a cleaning Outside of...

Страница 34: ...le A0013949 Channel 2 of a process variable A0013951 Channel 4 of a process variable Status symbols symbols for event level The status symbol pertains to a diagnostics event that is relevant to the di...

Страница 35: ...nu In the wizard Display symbol for wizard Omission symbol for operating menu levels in between Name of current Submenu Wizard Parameters Examples A0013973 Display A0013968 Display For more informatio...

Страница 36: ...rd A0013972 Parameters within a wizard No display symbol exists for parameters in submenus Lock symbols A0013963 Parameter locked When displayed in front of a parameter name indicates that the paramet...

Страница 37: ...ess keys simultaneously Exits the wizard and takes you to the next higher level General operating functions A0013953 Minus Enter key combination press and hold down the keys simultaneously Reduces the...

Страница 38: ...8 Deletes one character immediately to the left of the input position Operating elements General operating functions A0013969 Minus key In the input mask moves the selection bar to the left backwards...

Страница 39: ...Calling up and closing the context menu The user is in the measured value display 1 Press F for 2 s The context menu opens XXXXXXXXXX kg h 20 50 Setup Conf backup disp Simulation A0014003 EN 2 Press...

Страница 40: ...w with symbols and operating elements 35 Example Setting the number of displayed measured values to 2 values XXXX XXX X X 20 50 XXXX XXX X X 2 s 0104 1 Language 0091 1 0098 1 0098 1 0098 1 Language Di...

Страница 41: ...of the parameter and thus support fast and reliable commissioning Calling up and closing the help text The user is in the navigation view and the selection bar is on a parameter 1 Press F for 2 s The...

Страница 42: ...er 20 mA value to 20 kg s 0 1 2 3 5 6 7 8 9 4 20 kg s 0 1 2 3 5 6 7 8 9 4 100 000 0 1 2 3 5 6 7 8 9 4 100 000 0 1 2 3 5 6 7 8 9 4 0 1 2 3 5 6 7 8 9 4 0 1 2 3 5 6 7 8 9 4 2 0 1 2 3 5 6 7 8 9 4 2 kg s k...

Страница 43: ...n incorrect access code is entered the user obtains the access rights of the Operator role The user role with which the user is currently logged on is indicated by the Access status display parameter...

Страница 44: ...ng the S O F keys After enabling the keypad lock XXXXXXXXXX kg h 20 50 Keylock on A0014000 EN After disabling the keypad lock XXXXXXXXXX kg h 20 50 Keylock off A0014001 EN If the user attempts to acce...

Страница 45: ...iles See data 49 8 4 2 FieldCare Function scope FDT based plant asset management tool from Endress Hauser It can configure all smart field units in a system and help you manage them By using the statu...

Страница 46: ...e data 49 8 4 4 SIMATIC PDM Function scope SIMATIC PDM is a standardized manufacturer independent program from Siemens for the operation configuration maintenance and diagnosis of intelligent field de...

Страница 47: ...via HART protocol 1 PLC programmable logic controller 2 Transmitter power supply unit e g RN221N with communication resistor 3 Connection for Commubox FXA191 FXA195 and Field Communicator 375 475 4 Fi...

Страница 48: ...Proline Promass E TB2 HART 48 Endress Hauser Via CDI interface 1 2 A0014019 10 Operating options via CDI interface 1 Computer with operating tool FieldCare 2 Commubox FXA291 connected to the CDI inter...

Страница 49: ...description file for each individual operating tool with source Operating tools Operating tool Sources for obtaining device descriptions Field Xpert SFX100 Use update function of handheld terminal Fi...

Страница 50: ...w Volume flow Corrected volume flow Density Reference density Temperature Measured values for SV TV QV secondary tertiary and quarternary variable Mass flow Volume flow Corrected volume flow Density R...

Страница 51: ...vice After a successful startup the local display switches automatically from the startup display to the measured value display If nothing appears on the local display or a diagnostic message is displ...

Страница 52: ...vigation to the Setup menu XXXX XXX X X 20 50 Language 0104 1 Display operat Setup Diagnostic Display operat Setup Main menu English Main menu 1 2 3 4 Select medium Setup Curr output 2 Curr output 1 A...

Страница 53: ...the wizard Reference sound velocity Pressure compensation Temperature coefficient sound velocity Select medium Liquid Gas Select gas type Air Others Off Fixed Pressure value End of wizard A0013795 EN...

Страница 54: ...99999 m s 0 m s Temperature coefficient sound velocity Enter the temperature coefficient of the sound velocity of the gas Max 15 digit positive floating point number 0 m s K Pressure compensation Enab...

Страница 55: ...nu Current output 1 Setup menu Current output 2 Sequence of the wizard Part 1 corresponding part 2 next page Assign current output Off Mass flow Volume flow Density Temperature Mass flow unit Volume f...

Страница 56: ...t Fixed reference density Corrected volume flow Reference density Linear expansion coefficient Square expansion coefficient Temperature unit Reference temperature Reference density unit 20 mA value Fa...

Страница 57: ...the unit for density Result The selected unit applies for Current outputs Low value partial filled pipe detection High value partial filled pipe detection Simulation process variable Density adjustme...

Страница 58: ...ariable Unit picklist Country dependent C Celsius F Fahrenheit Reference temperature Enter the reference temperature for calculating the reference density Max 15 digit floating point number with sign...

Страница 59: ...splay Value 1 display 0 bargraph value 1 100 bargraph value 1 Value 2 display Value 3 display 0 bargraph value 3 100 bargraph value 3 Value 4 display End of wizard A0013797 EN 14 Display wizard in the...

Страница 60: ...h sign Depends on country and nominal diameter Value 2 display Select the measured value shown on the local display None Mass flow Volume flow Corrected volume flow Density Reference density Temperatu...

Страница 61: ...ne Mass flow Volume flow Corrected volume flow Density Reference density Temperature Totalizer 1 Totalizer 2 Totalizer 3 Current output 1 Current output 2 None 10 4 4 Configuring the output conditioni...

Страница 62: ...s variable Pressure shock suppression On value low flow cutoff Off value low flow cutoff Operating mode totalizer 1 Slight Moderate Strong Display damping Display damping Display damping Damping outpu...

Страница 63: ...fluctuations in the measured value caused by process conditions 0 to 999 s 1 s Measuring mode output 1 2 Select the measuring mode for current output Forward flow Forward reverse flow Reverse flow com...

Страница 64: ...A0013799 EN 16 Low flow cut off wizard in the Setup menu Parameter overview with brief description Parameters Description Selection User entry Factory setting Assign process variable Select the proce...

Страница 65: ...eter overview with brief description Parameters Description Selection User entry Factory setting Assign process variable Select a process variable to detect empty or partially filled pipes Off Density...

Страница 66: ...Display operat Setup Diagnostic Display operat Setup Main menu English Main menu 1 2 3 4 Select medium Setup Curr output 2 Curr output 1 Low flow cut off Setup Advanced setup Partial pipe det 5 Ent a...

Страница 67: ...or special characters e g In the factory configuration the tag name appears at the top left in the header 1 XXXXXXXXX A0013375 1 Device tag The number of characters displayed depends on the character...

Страница 68: ...riable Unit picklist Country dependent l h gal min us Volume Select the unit for volume Result The selected unit is taken from Volume flow unit Country dependent l gal us Corrected volume flow unit Se...

Страница 69: ...us F Fahrenheit Length unit Select the unit of length for the nominal diameter Unit picklist Country dependent mm in Pressure unit Select the unit for pipe pressure Unit picklist Country dependent bar...

Страница 70: ...es the picklist of the Unit parameter Off Mass flow Volume flow Corrected volume flow Mass flow Unit One of the following options is selected in the Assign process variable parameter Mass flow Volume...

Страница 71: ...0 bargraph value 1 59 Decimal places 1 Value 2 display 59 Decimal places 2 Value 3 display 59 0 bargraph value 3 59 100 bargraph value 3 59 Decimal places 3 Value 4 display 59 Decimal places 4 Display...

Страница 72: ...special characters Separator Select the decimal separator for numerical values point comma point 10 6 Configuration management After commissioning you can save the current device configuration copy it...

Страница 73: ...ation device alarm Parameter overview with brief description Parameters Prerequisite Description Selection User entry Factory setting Assign simulation process variable Select a process variable for t...

Страница 74: ...e symbol appears in front of all write protected parameters Parameters that can always be changed The write protection does not include certain parameters that do not affect the measurement Despite th...

Страница 75: ...sening the securing clamp and unscrewing the electronics compartment cover 2 Pull out the display module with a gentle rotation movement 3 A0013909 To make it easier to access the lock switch attach t...

Страница 76: ...f the measured value display and in the navigation view in front of the parameters 5 Feed the spiral cable into the gap between the housing and main electronics module and plug the display module into...

Страница 77: ...lues You can read all measured values using the Measured value menu Navigation path Diagnostics Measured val The following three submenus are available Process variable submenu Process variab Mass flo...

Страница 78: ...et all totalizers Navigation path Display operat menu Operation Operation submenu Operation Control totalizer 1 Preset value 1 Control totalizer 2 Preset value 2 Control totalizer 3 Preset value 3 Res...

Страница 79: ...ta logging submenu Data logging Assign channel 1 Assign channel 2 Assign channel 3 Assign channel 4 Logging interval Clear logging data Display channel 1 Display channel 2 Display channel 3 Display ch...

Страница 80: ...ration Proline Promass E TB2 HART 80 Endress Hauser If the length of the logging interval or the assignment of the process variables to the channels is changed the content of the data logging is delet...

Страница 81: ...e Order spare part 91 I O electronics module is defective Order spare part 91 Device shows correct value on local display but signal output is incorrect though in the valid current range Configuration...

Страница 82: ...ng elements Status signals A0013956 Failure A device error is present The measured value is no longer valid A0013957 Maintenance required Maintenance is required The measured value is still valid A001...

Страница 83: ...ional pending diagnostic messages can be shown in the Diagnose list submenu 84 Past diagnostic messages that are no longer pending are shown in the Event logbook submenu 88 Operating elements Operatin...

Страница 84: ...07 Failure F Maintenance required M Function check C Out of specification S A0014014 EN Calling up remedial measures 1 Navigate to the Diagnostics menu In the Actual diagnostics parameter the diagnost...

Страница 85: ...ert menu System Diagn handling Event level Event no xxx 0723 1 Event no 044 Event level Event no 801 Event no 274 Warning A0014048 EN You can assign the following options to the diagnostics event as t...

Страница 86: ...les incompatible 1 Check electronic modules 2 Change I O or main electronics module Alarm 1 F261 Electr modules 1 Restart device 2 Check electronic modules 3 Change I O Modul or main electronics Alarm...

Страница 87: ...ulation Warning 12 5 5 Diagnostics for the process Diagnostics event Event text Remedial measures Event level from the factory S801 Supply voltage Supply voltage too low Increase voltage Warning 1 M80...

Страница 88: ...gas or solid content 1 Check process conditions 2 Increase supply 3 Check main electronic or sensor Warning 1 1 Event level can be changed Chapter 12 5 1 Adapting event level 12 6 Event logbook 12 6 1...

Страница 89: ...intenance required M Function check C Out of specification S Information I 12 6 3 Overview of information events Unlike a diagnostic event an information event is displayed in the event logbook only a...

Страница 90: ...hooting Proline Promass E TB2 HART 90 Endress Hauser Information event Event text I1221 Error during zero point adjustment I1222 Zero point adjustment OK I1227 Sensor emergency mode enabled I1228 Sens...

Страница 91: ...umentation XA and certificates Document every repair and each conversion and enter them into the W M life cycle management database 13 2 Spare parts A few interchangeable measuring device components a...

Страница 92: ...Repair Proline Promass E TB2 HART 92 Endress Hauser 13 3 Endress Hauser services For information about service and spare parts contact your Endress Hauser distributor...

Страница 93: ...rve the maximum permitted fluid temperature for the measuring device 103 14 2 Measuring and test equipment Endress Hauser offers a wide variety of measuring and test equipment such as W M or device te...

Страница 94: ...subsequently from Endress Hauser are available for the device Detailed information on the order code in question is available from your local Endress Hauser sales center or on the product page of the...

Страница 95: ...zardous to health e g flammable toxic caustic or carcinogenic Costs incurred for waste disposal and injury due to inadequate cleaning will be charged to the owner operator Complete all fields of the D...

Страница 96: ...device and Connecting the measuring device in the logically reverse sequence Observe the safety instructions 17 2 Disposing of the measuring device WARNING Danger to personnel and environment from flu...

Страница 97: ...ple Measuring system The device consists of a transmitter and a sensor One device version is available Compact version transmitter and sensor form a mechanical unit For information on the structure of...

Страница 98: ...03 Operable flow range Greater than 1000 1 Flow rates above the preset full scale value do not overload the amplifier so the totalized values are registered correctly 18 4 Output Output signal 2 wire...

Страница 99: ...o 660 mW Operation with current output 1 and 2 130 to 1320 mW Power supply failure Totalizers stop at the last value measured Configuration is retained in the HistoROM Error messages incl total operat...

Страница 100: ...easured error In addition to the values indicated the measured error at the current output is typically 4 mA o r of reading 1 g cm3 1 kg l T fluid temperature Mass flow and volume flow liquids 0 25 o...

Страница 101: ...0 1 72 2 646 3 100 1 180 6 615 1 2 25 1 720 26 46 0 3 10 1 1800 66 15 0 25 2 1 9000 330 75 0 25 Repeatability o r of reading 1 g cm3 1 kg l T fluid temperature Mass flow and volume flow liquids 0 125...

Страница 102: ...between calibration pressure and process pressure o r of reading DN o r bar mm in 8 3 8 no influence 15 no influence 25 1 no influence 40 1 no influence 50 2 0 009 Design fundamentals Dependent on the...

Страница 103: ...nsor 40 to 140 C 40 to 284 F Seals No internal seals Medium density 0 1 to 2000 kg m3 0 2 to 4410 lb m3 Medium pressure range nominal pressure An overview of the material load diagrams pressure temper...

Страница 104: ...sizing tool 94 Pressure loss in SI units Pressure loss depends on the fluid properties and on the flow rate The following formulas can be used to approximately calculate the pressure loss Reynolds nu...

Страница 105: ...the device see the Technical Information document Mechanical construction section Rupture disk optional Rupture disks cannot be combined with the separately available heating jacket 94 Weight Weight i...

Страница 106: ...nless steel 1 4404 316L For all process connections except flanges as per JIS B2220 Stainless steel SUS 316L For flanges as per JIS B2220 List of all available process connections 107 Measuring tubes...

Страница 107: ...11864 1 Form A DIN 11864 2 Form A flat flange with groove For information on the materials of the process connections 106 18 11 Human interface Local display SD02 Display elements 4 line display Form...

Страница 108: ...parate Safety Instructions XA document Reference is made to this document on the nameplate Hygienic compatibility 3A approval Other standards and guidelines EN 60529 Degrees of protection by housing I...

Страница 109: ...pecified in Annex I of the Pressure Equipment Directive 97 23 EC Devices not bearing this marking PED are designed and manufactured according to good engineering practice They meet the requirements of...

Страница 110: ...e measuring device 52 Communication specific data 49 Connecting cable Requirements 25 Connecting the measuring device 25 Connection see Electrical connection Connection preparations 25 Connection tool...

Страница 111: ...elease date 49 Version 49 Flow direction 18 22 Flow limit 103 Fluids 8 Function check 51 Function scope AMS Device Manager 46 Field Communicator 46 Field Communicator 475 46 Field Xpert 45 FieldCare 4...

Страница 112: ...g device configuration 52 For specific settings 66 Mounting 17 Mounting dimensions see Installation dimensions Mounting location 17 Mounting preparations 22 Mounting requirements Inlet and outlet runs...

Страница 113: ...instructions 8 Safety symbols 6 Safety icons see Safety Seals Medium temperature range 103 Secondary containment pressure range 103 Sensor Medium temperature range 103 Mounting 22 Sensor heating 20 Se...

Страница 114: ...Terminals 99 Text editor 37 Thermal insulation 20 Tool tip see Help text Tools For electrical connection 25 For mounting 22 Transport 15 Transmitter Connecting the signal cables 26 Turning the displa...

Страница 115: ..._____________________________________________________________________________________ We hereby certify that this declaration is filled out truthfully and completely to the best of our knowledge We fu...

Страница 116: ...www endress com worldwide 71112405 71112405 BA01027D 06 EN 06 10 71112405 EH COSIMA ProMoDo...