Modbus RS485 Register Information
Proline Promag 300 Modbus RS485
244
Hauser
"Process param." submenu
Navigation: Expert → Sensor → Process param.
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Filter options (6710)
2273
Integer
Read / Write
1 = Adaptive
2 = Adaptive CIP on
3 = Dynamic
4 = Dynamic CIP on
5 = Binomial
6 = Binomial CIP on
Flow damping (6661)
2274
Integer
Read / Write
0 to 15
Flow override (1839)
5503
Integer
Read / Write
0 = Off
1 = On
Conduct. damping (1803)
5508 to 5509
Float
Read / Write
0 to 999.9 s
Temp. damping (1886)
2483 to 2484
Float
Read / Write
0 to 999.9 s
Conduct. measur. (6514)
2268
Integer
Read / Write
0 = Off
1 = On
Cond. temp.coeff (1891)
2886 to 2887
Float
Read / Write
Signed floating-point number
Ref.density (1885)
2536 to 2537
Float
Read / Write
Positive floating-point number
"Low flow cut off" submenu
Navigation: Expert → Sensor → Process param. → Low flow cut off
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Assign variable (1837)
5101
Integer
Read / Write
0 = Off
1 = Volume flow
2 = Mass flow
3 = Correct.vol.flow
On value (1805)
5138 to 5139
Float
Read / Write
Positive floating-point number
Off value (1804)
5104 to 5105
Float
Read / Write
0 to 100.0 %
Pres. shock sup. (1806)
5140 to 5141
Float
Read / Write
0 to 100 s
"Empty pipe det." submenu
Navigation: Expert → Sensor → Process param. → Empty pipe det.
Parameter
Register
Data type
Access
Selection / User entry / User interface
→
Empty pipe det. (1860)
5106
Integer
Read / Write
0 = Off
1 = On
Switch point EPD (6562)
2890 to 2891
Float
Read / Write
0 to 100 %
Response time (1859)
5108 to 5109
Float
Read / Write
0 to 100 s
New adjustment (6560)
2335
Integer
Read / Write
0 = Cancel
1 = Empty pipe adj.
2 = Full pipe adjust
Progress (6571)
2336
Integer
Read
0 = Not ok
6 = Ok
8 = Busy
Empty pipe value (6527)
2181 to 2182
Float
Read
Positive floating-point number
Full pipe value (6548)
2832 to 2833
Float
Read
Positive floating-point number
Meas. value EPD (6559)
2298 to 2299
Float
Read
Positive floating-point number