Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
79
"Empty pipe det." submenu
Navigation
Expert → Sensor → Process param. → Empty pipe det.
‣
Empty pipe det.
Empty pipe det. (1860)
Switch point EPD (6562)
Response time (1859)
New adjustment (6560)
Progress (6571)
Empty pipe value (6527)
Full pipe value (6548)
Meas. value EPD (6559)
Empty pipe det.
Navigation
Expert → Sensor → Process param. → Empty pipe det. → Empty pipe det. (1860)
Description
Use this function to switch empty pipe detection on and off.
Selection
• Off
• On
Factory setting
Off
Switch point EPD
Navigation
Expert → Sensor → Process param. → Empty pipe det. → Switch point EPD (6562)
Prerequisite
The
On
option is selected in the
Empty pipe det.
Description
Use this function to enter the percentage threshold value of the resistance in relation to
the adjustment values.
User entry
0 to 100 %
Factory setting
10 %