Description of device parameters
Proline Promag 300 Modbus RS485
80
Hauser
Response time
Navigation
Expert → Sensor → Process param. → Empty pipe det. → Response time (1859)
Prerequisite
In the
Empty pipe det.
parameter (→ 79), the
On
option is selected.
Description
Use this function to enter the minimum length of time (debouncing time) the signal must
be present for the diagnostic message
S862 Pipe empty
to be triggered if the measuring
pipe is empty or partially full.
User entry
0 to 100 s
Factory setting
1 s
New adjustment
Navigation
Expert → Sensor → Process param. → Empty pipe det. → New adjustment (6560)
Prerequisite
The
On
option is selected in the
Empty pipe det.
Description
For selecting whether to perform an empty pipe or full pipe adjustment.
Selection
• Cancel
• Empty pipe adj.
• Full pipe adjust
Factory setting
Cancel
Progress
Navigation
Expert → Sensor → Process param. → Empty pipe det. → Progress (6571)
Prerequisite
The
On
option is selected in the
Empty pipe det.
Description
Use this function to view the progress.
User interface
• Ok
• Busy
• Not ok