Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
147
The lower the pulse value, the
• better the resolution.
• the higher the frequency of the pulse response.
Pulse width
Navigation
Expert → Output → Double pulse out → Pulse width (0986)
Description
Use this function to enter the duration of the output pulse.
User entry
0.5 to 2
000 ms
Factory setting
0.5 ms
Additional information
For a detailed description and example:
Pulse width
Phase shift
Navigation
Expert → Output → Double pulse out → Phase shift (0992)
Description
Use this function to select the degree of phase shift.
Selection
• 90°
• 180°
Factory setting
90°
Additional information
Selection
• 90°
Phase shift by a quarter period.
• 180°
Phase shift by a half period, which is equivalent to a phase reversal.
Measuring mode
Navigation
Expert → Output → Double pulse out → Measuring mode (0984)
Description
Use this function to select the measuring mode for the double pulse output.
Selection
• Forward flow
• Forward/Reverse
• Reverse flow
• Rev. flow comp.
Factory setting
Forward flow