Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
91
Additional information
Description
Corrected value = (factor × value) + offset
Mass flow offset
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow offset (1841)
Description
Use this function to enter the zero point shift for the mass flow trim. The mass flow unit
on which the shift is based is kg/s.
User entry
Signed floating-point number
Factory setting
0 kg/s
Additional information
Description
Corrected value = (factor × value) + offset
Mass flow factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Mass flow factor (1846)
Description
Use this function to enter a quantity factor (without time) for the mass flow. This
multiplication factor is applied over the mass flow range.
User entry
Positive floating-point number
Factory setting
1
Additional information
Description
Corrected value = (factor × value) + offset
Conduct. offset
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Conduct. offset (1848)
Prerequisite
The
On
option is selected in the
Conduct. measur.
parameter (→ 74) parameter.
Description
Use this function to enter the zero point shift for the conductivity trim. The conductivity
unit on which the shift is based is S/m.
User entry
Signed floating-point number
Factory setting
0 S/m