Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
137
Switch-off delay
Navigation
Expert → Output → PFS output 1 to n → Switch-off delay (0465–1 to n)
Prerequisite
• The
Switch
option is selected in the
Operating mode
parameter (→ 120).
• The
Limit
option is selected in the
Switch out funct
Description
Use this function to enter a delay time for switching off the switch output.
User entry
0.0 to 100.0 s
Factory setting
0.0 s
Failure mode
Navigation
Expert → Output → PFS output 1 to n → Failure mode (0486–1 to n)
Description
Use this function to select a failsafe mode for the switch output in the event of a device
alarm.
Selection
• Actual status
• Open
• Closed
Factory setting
Open
Additional information
Options
• Actual status
In the event of a device alarm, faults are ignored and the current behavior of the input
value is output by the switch output. The
Actual status
option behaves in the same way
as the current input value.
• Open
In the event of a device alarm, the switch output's transistor is set to
non-conductive
.
• Closed
In the event of a device alarm, the switch output's transistor is set to
conductive
.
Switch status 1 to n
Navigation
Expert → Output → PFS output 1 to n → Switch status 1 to n (0461–1 to n)
Prerequisite
The
Switch
option is selected in the
Operating mode
Description
Displays the current switch status of the status output.
User interface
• Open
• Closed