Description of device parameters
Proline Promag 300 Modbus RS485
120
Hauser
Signal mode
Navigation
Expert → Output → PFS output 1 to n → Signal mode (0490–1 to n)
Description
Use this function to select the signal mode for the pulse/frequency/switch output.
Selection
• Passive
• Active
Factory setting
Passive
Operating mode
Navigation
Expert → Output → PFS output 1 to n → Operating mode (0469–1 to n)
Description
Use this function to select the operating mode of the output as a pulse, frequency or switch
output.
Selection
• Pulse
• Frequency
• Switch
Factory setting
Pulse
Additional information
"Pulse" option
Quantity-dependent pulse with configurable pulse width
• Whenever a specific mass, volume or corrected volume is reached (pulse value), a pulse
is output, the duration of which was set previously (pulse width).
• The pulses are never shorter than the set duration.
Example
• Flow rate approx. 100 g/s
• Pulse value 0.1 g
• Pulse width 0.05 ms
• Pulse rate 1
000 Impuls/s