Description of device parameters
Proline Promag 300 Modbus RS485
170
Hauser
Options
Description
totalize
The totalizer is set to the defined start value from the
Preset value
parameter and
the totaling process is restarted.
Hold
Totalizing is stopped.
Preset value 1 to n
Navigation
Expert → Application → Totalizer 1 to n → Preset value 1 to n (0913–1 to n)
Prerequisite
One of the following options is selected in the
Assign variable
the
Totalizer 1 to n
submenu:
Volume flow
Description
Use this function to enter a start value for the Totalizer 1 to n.
User entry
Signed floating-point number
Factory setting
0 l
Additional information
Entry
The unit of the selected process variable is specified for the totalizer in the
Unit
totalizer
parameter (→ 168).
Example
This configuration is suitable for applications such as iterative filling processes with a fixed
batch quantity.
Failure mode
Navigation
Expert → Application → Totalizer 1 to n → Failure mode (0901–1 to n)
Description
Use this function to select how a totalizer behaves in the event of a device alarm.
Selection
• Stop
• Actual value
• Last valid value
Factory setting
Stop