Description of device parameters
Proline Promag 300 Modbus RS485
74
Hauser
Conduct. damping
Navigation
Expert → Sensor → Process param. → Conduct. damping (1803)
Prerequisite
In the
Conduct. measur.
parameter (→ 74), the
On
option is selected.
Description
Use this function to enter a time constant for conductivity damping (PT1 element).
User entry
0 to 999.9 s
Factory setting
0 s
Additional information
Description
The damping is performed by a PT1 element
.
User entry
• Value = 0: no damping
• Value > 0: damping is increased
Damping is switched off if
0
is entered (factory setting).
Temp. damping
Navigation
Expert → Sensor → Process param. → Temp. damping (1886)
Prerequisite
One of the following conditions is met:
• Order code for "Sensor option", option
CI
"Medium temperature measurement"
or
• The temperature is read into the flowmeter from an external device.
Description
Use this function to enter the time constant for temperature damping.
User entry
0 to 999.9 s
Factory setting
0 s
Conduct. measur.
Navigation
Expert → Sensor → Process param. → Conduct. measur. (6514)
Prerequisite
The
On
option is selected in the
Conduct. measur.
parameter (→ 74) parameter.
Description
Use this function to enable and disable conductivity measurement.
2)
Proportional behavior with first-order lag