Proline Cubemass C 100 HART
Operation
Hauser
79
11
Operation
11.1 Reading device locking status
The write protection types that are currently active can be determined using the
Locking
status
parameter.
Navigation
"Operation" menu → Locking status
Function scope of "Locking status" parameter
Options
Description
Hardware locked
The locking switch (DIPswitch) for locking the hardware is activated on the main
electronic module. This prevents write access to the parameters → 78.
Temporarily locked
Due to internal processing in the device (e.g. up-/downloading of data, reset), write
access to the parameters is blocked for a short time. Once the internal processing
has been completed, the parameters can be changed once again.
11.2 Reading measured values
Using the
Measured values
submenu , it is possible to read all the measured values.
Diagnostics → Measured values
11.2.1 Process variables
The
Process variables
submenu contains all the parameters needed to display the current
measured values for every process variable.
Navigation
"Diagnostics" menu → Measured values → Process variables
Process variables
Mass flow
Volume flow
Corrected volume flow
Density
Reference density
Temperature
Pressure value
Parameter overview with brief description
Parameter
Description
User interface
Factory setting
Mass flow
Displays the mass flow currently measured.
Signed floating-point number
–
Volume flow
Displays the calculated volume flow.
Signed floating-point number
–
Corrected volume flow
Displays the corrected volume flow currently
calculated.
Signed floating-point number
–