Hauser
Parameter overview with brief description
Parameter
Description
Selection / User interface
Factory setting
Installation direction
Set sign of flow direction to match the
direction of the arrow on the sensor.
• Flow in arrow direction
• Flow against arrow direction
Flow in arrow direction
Zero point adjustment control
Start zero point adjustment.
• Cancel
• Busy
• Zero point adjust failure
• Start
Cancel
Progress
Shows the progress of the process.
0 to 100 %
0 %
10.3.3 Configuring the totalizer
In the
"Totalizer 1 to n" submenu
the individual totalizers can be configured.
Navigation
"Setup" menu → Advanced setup → Totalizer 1 to n
Totalizer 1 to n
→
Assign process variable
Unit totalizer
Failure mode
Parameter overview with brief description
Parameter
Description
Selection
Factory setting
Assign process variable
Select process variable for totalizer.
• Off
• Volume flow
• Mass flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
Mass flow
Unit totalizer
Select process variable totalizer unit.
Unit choose list
kg
Totalizer operation mode
Select totalizer calculation mode.
• Net flow total
• Forward flow total
• Reverse flow total
Net flow total
Failure mode
Define totalizer behavior in alarm condition. • Stop
• Actual value
• Last valid value
Stop