Proline Cubemass C 100 HART
Commissioning
Hauser
65
Navigation
"Expert" menu → Communication → HART input → Configuration
HART input
→
Capture mode
Device ID
Device type
Manufacturer ID
Burst command
Slot number
Timeout
Failure mode
Failure value
Parameter overview with brief description
Parameter
Description
Selection / User entry
Factory setting
Capture mode
Select capture mode via burst or master
communication.
• Off
• Burst network
• Master network
Off
Manufacturer ID
Enter manufacture ID of external device.
0 to 255
0
Device ID
Enter device ID of external device.
Positive integer
0
Device type
Enter device type of external device.
0 to 255
0
Burst command
Select command to read in external process
variable.
• Command 1
• Command 3
• Command 9
• Command 33
Command 1
Slot number
Define position of external process variable
in burst command.
1 to 4
1
Timeout
Enter deadline for process variable of
external device.
If the deadline is exceeded, diagnostic
message
F410 data transmission
is
output.
1 to 120 s
5 s
Failure mode
Define behavior if external process variable
is missed.
• Alarm
• Last valid value
• Defined value
Alarm
Failure value
Enter value to be used by the device if input
value from external device is missing.
Signed floating-point number
0
10.2.8 Configuring the output conditioning
The
Output conditioning
wizard contains all the parameters that must be configured for
the configuration of output conditioning.