Diagnostics and troubleshooting
Proline Cubemass C 100 HART
84
Hauser
12.2 Diagnostic information via light emitting diodes
12.2.1 Transmitter
Various light emitting diodes (LEDs) on the main electronics module of the transmitter
provide information on device status.
LED
Color
Meaning
Power
Off
Supply voltage is off or too low
Green
Supply voltage is ok
Link/Activity
Orange
Link available but no activity
Flashing orange
Activity present
Communication
Flashing white
HART communication is active.
12.3 Diagnostic information in FieldCare
12.3.1 Diagnostic options
Any faults detected by the measuring device are displayed on the home page of the
operating tool once the connection has been established.
Xxxxxx/…/…/
Xxxxxx
P
P
P
+
+
+
+
Diagnostics 1:
Remedy information:
Failure (F)
Function check (C)
Out of spezification (S)
Maintenance required (M)
C485 Simulation measured vari...
Deactivate Simulation (Service...
Diagnostics 1:
C485 Simu...
Remedy information:
Deactivate...
Access status tooling:
Mainenance
Operation
Setup
Diagnostics
Expert
Function check (C)
Status signal:
Device tag:
Device name:
Mass flow:
Volume flow:
Xxxxxxx
12.34
kg/h
12.34
m /h
³
Xxxxxxx
1
2
3
A0021799-EN
1
Status area with status signal
2
3
Remedial measures with Service ID
Furthermore, diagnostic events that have occurred can be viewed in the
Diagnostics
menu:
Status signals
The status signals provide information on the state and reliability of the device by
categorizing the cause of the diagnostic information (diagnostic event).
Содержание Proline Cubemass C 100 HART
Страница 126: ...www addresses endress com...