Operation options
Proline Cubemass C 100 HART
40
Hauser
Network connections
Only the active network connections to the measuring device should be used.
Switch off all other network connections.
In the event of connection problems:
Measuring device: Via CDI-RJ45 service interface
Device
CDI-RJ45 service interface
Measuring device
The measuring device has an RJ45 interface.
Web server
Web server must be enabled; factory setting: ON
For information on enabling the Web server → 43
8.4.3
Establishing a connection
Via service interface (CDI-RJ45)
Preparing the measuring device
Configuring the Internet protocol of the computer
The following information refers to the default Ethernet settings of the device.
IP address of the device: 192.168.1.212 (factory setting)
1. Switch on the measuring device.
2. Connect to the computer using a cable → 116.
3. If a 2nd network card is not used, close all the applications on the notebook.
Applications requiring Internet or a network, such as e-mail, SAP applications,
Internet or Windows Explorer.
4. Close any open Internet browsers.
5. Configure the properties of the Internet protocol (TCP/IP) as defined in the table:
IP address
192.168.1.XXX; for XXX all numerical sequences except: 0, 212 and 255 → e.g.
192.168.1.213
Subnet mask
255.255.255.0
Default gateway
192.168.1.212 or leave cells empty
Starting the Web browser
1. Start the Web browser on the computer.