Commissioning
Proline Cubemass C 100 HART
76
Hauser
Value current output
Frequency simulation
Frequency value
Pulse simulation
Pulse value
Switch output simulation
Switch status
Simulation device alarm
Simulation diagnostic event
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Assign simulation process variable
–
Select a process variable for
the simulation process that is
activated.
• Off
• Mass flow
• Volume flow
• Corrected volume
flow
• Density
• Reference density
• Temperature
• Dynamic viscosity
• Kinematic viscosity
• Temp.
compensated
dynamic viscosity
• Temp.
compensated
kinematic viscosity
• Concentration
• Target mass flow
• Carrier mass flow
Off
Value process variable
A process variable is selected
in the
Assign simulation
process variable
parameter.
Enter the simulation value for
the selected process variable.
Signed floating-point
number
0
Simulation current output 1
–
Switch simulation of the
current output on and off.
• Off
• On
Off
Value current output 1
The
On
option is selected in
the
Current output
simulation
parameter.
Enter the current value for
simulation.
3.59 · 10
-3
to 22.5 · 1
0
-3
mA
3.59 mA
Frequency simulation 1
–
Switch simulation of the
frequency output on and off.
• Off
• On
Off
Frequency value 1
The
On
option is selected in
the
Frequency output
simulation
parameter.
Enter the frequency value for
simulation.
0.0 to 12
500.0 Hz
0.0 Hz
Pulse simulation 1
The
Down-count. val.
option
is selected in the
Simulation
pulse output
parameter.
Switch simulation of the pulse
output on and off.
If the
Fixed value
option
is selected, the
Pulse
width
parameter defines
the pulse width of the
pulses output.
• Off
• Fixed value
• Down-counting
value
Off
Содержание Proline Cubemass C 100 HART