Operation
Proline Cubemass C 100 HART
80
Hauser
Parameter
Description
User interface
Factory setting
Density
Displays the density currently measured.
Signed floating-point number
–
Reference density
Displays the reference density currently
calculated.
Signed floating-point number
–
Temperature
Displays the temperature currently
measured.
Signed floating-point number
Pressure value
Displays either a fixed or external pressure
value.
Signed floating-point number
11.2.2 Totalizer
The
"Totalizer" submenu
contains all the parameters needed to display the current
measured values for every totalizer.
Navigation
"Diagnostics" menu → Measured values → Totalizer
Structure of the submenu
Totalizer
→
Totalizer value
Totalizer overflow
Parameter overview with brief description
Parameter
Prerequisite
Description
User interface
Factory setting
Totalizer value 1 to n
In the
Assign process variable
parameter of
Totalizer 1 to n
submenu one of the following
options is selected:
• Volume flow
• Mass flow
• Corrected volume flow
Displays the current totalizer
counter value.
Signed floating-point
number
0 kg
Totalizer overflow 1 to n
In the
Assign process variable
parameter of
Totalizer 1 to n
submenu one of the following
options is selected:
• Volume flow
• Mass flow
• Corrected volume flow
Displays the current totalizer
overflow.
–
32
000.0 to 32
000.0
0
11.2.3 Output values
The
"Output values" submenu
contains all the parameters needed to display the current
measured values for every output.
Navigation
"Diagnostics" menu → Measured values → Output values
Structure of the submenu
Output values
→
Output current 1
Measured current 1