Proline Cubemass C 100 HART
Technical data
Hauser
107
SI units
DN
1:1
1:10
1:20
1:50
1:100
1:500
[mm]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
1
20
2
1
0.4
0.2
0.04
2
100
10
5
2
1
0.2
4
450
45
22.5
9
4.5
0.9
6
1
000
100
50
20
10
2
US units
DN
1:1
1:10
1:20
1:50
1:100
1:500
[inch]
[lb/min]
[lb/min]
[lb/min]
[lb/min]
[lb/min]
[lb/min]
¹⁄₂₄
0.735
0.074
0.037
0.015
0.007
0.001
¹⁄₁₂
3.675
0.368
0.184
0.074
0.037
0.007
¹⁄₈
16.54
1.654
0.827
0.331
0.165
0.033
¹⁄₄
36.75
3.675
1.838
0.735
0.368
0.074
Accuracy of outputs
The output accuracy must be factored into the measured error if analog outputs are
used, but can be ignored for fieldbus outputs (e.g. Modbus RS485, EtherNet/IP).
The outputs have the following base accuracy specifications.
Current output
Accuracy
Max. ±5 µA
Pulse/frequency output
o.r. = of reading
Accuracy
Max. ±50 ppm o.r. (over the entire ambient temperature range)
Repeatability
o.r. = of reading; 1 g/cm
3
= 1 kg/l; T = medium temperature
Base repeatability
Mass flow and volume flow (liquids)
±0.05 % o.r.
Mass flow (gases)
±0.25 % o.r.
Density (liquids)
±0.00025 g/cm
3
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)