Proline Cubemass C 100 HART
Commissioning
Hauser
71
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Corrected volume flow calculation
–
Select reference density for
calculating the corrected
volume flow.
• Fixed reference
density
• Calculated
reference density
• Reference density
by API table 53
Calculated reference
density
External reference density
–
Shows external reference
density.
Floating point
number with sign
0 kg/Nl
Fixed reference density
The following option is
selected in the
Corrected
volume flow calculation
parameter:
Fixed reference density
Enter fixed value for reference
density.
Positive floating-
point number
1 kg/Nl
Reference temperature
The following option is
selected in the
Corrected
volume flow calculation
parameter:
Calculated reference density
Enter reference temperature
for calculating the reference
density.
–273.15 to 99
999 °C 20 °C
Linear expansion coefficient
The following option is
selected in the
Corrected
volume flow calculation
parameter:
Calculated reference density
Enter linear, medium-specific
expansion coefficient for
calculating the reference
density.
Signed floating-point
number
0.0
Square expansion coefficient
–
For media with a non-linear
expansion pattern: enter the
quadratic, medium-specific
expansion coefficient for
calculating the reference
density.
Signed floating-point
number
0.0
10.3.2 Carrying out a sensor adjustment
The
Sensor adjustment
submenu contains parameters that pertain to the functionality of
the sensor.
Navigation
"Setup" menu → Advanced setup → Sensor adjustment
Structure of the submenu
Sensor adjustment
→
Installation direction
→
Zero point adjustment
→
Zero point adjustment control
Progress