System integration
Hauser
49
Dynamic variables
Measured variables (HART device variables)
Tertiary dynamic variable (TV)
Density
Quaternary dynamic variable (QV)
Temperature
The assignment of the measured variables to the dynamic variables can be modified and
assigned as desired via the operating tool using the following parameters:
• Expert → Communication → HART output → Output → Assign PV
• Expert → Communication → HART output → Output → Assign SV
• Expert → Communication → HART output → Output → Assign TV
• Expert → Communication → HART output → Output → Assign QV
The following measured variables can be assigned to the dynamic variables:
Measured variables for PV (primary dynamic variable)
• Off
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Carrier pipe temperature
• Electronic temperature
• Oscillation frequency 0
• Frequency fluctuation 0
• Oscillation damping 0
• Oscillation damping fluctuation 0
• Signal asymmetry
• Exciter current 0
Measured variables for SV, TV, QV (secondary, tertiary and quaternary dynamic
variable)
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Electronic temperature
• Oscillation frequency
• Oscillation amplitude
• Oscillation damping
• Signal asymmetry
• External pressure
• Totalizer 1…3
9.2.1
Device variables
Device variables are permanently assigned. A maximum of eight device variables can be
transmitted.
Assignment
Device variables
0
Mass flow
1
Volume flow
2
Corrected volume flow
3
Density
4
Reference density