Table 2-1: Troubleshooting
(continued)
Error
Recommended action(s)
Cannot communicate with Field
Communicator
Refer to the Emerson 475 Field Communication User’s Manual,
Rev D. This manual may be downloaded from the following
location:
http://www2.emersonprocess.com/en-US/brands/
Field-Communicator/Pages/Documentation.aspx
Note
The 375 Field Communicator is no longer available for purchase
since the release of the 475 Field Communicator. Customer
support for the 375 Field Communicator remains available.
Cannot communicate with AMS
Device Manager
Refer to the AMS help documentation and support at the
following web site:
http://www.emersonprocess.com/ams/
Connect to multiple meters via
Ethernet when they are on the
same LAN
•
Configure each meter with a unique user-specified IP
address.
•
Contact your IT department for valid IP addresses for your
LAN and Gateway addresses.
•
Disable the DHCP server.
Connect to multiple meters via
Ethernet when they are on the
same hub but not connected to
an intranet LAN
•
Configure each meter with a unique user-specified IP
address.
•
Assign each meter on the hub a unique IP address within
the range 192.168.135.150 through 192.168.135.254
(Gateway address for each meter may be left unconfigured
as 0.0.0.0).
•
A PC may receive its IP address from an external DHCP
server; in this case, one and only one meter must have its
DHCP server enabled (the DHCP server will serve up to 10 IP
addresses to PCs attempting to talk to all meters on the
hub).
•
Once a meter's IP address is configured, the meter may be
connected to the hub and accessed using that IP address.
Configuration changed
One or more parameters have been modified in the meter's
configuration
•
Collect an Audit log using MeterLink in order to see what
configuration parameters changed and when they changed.
•
Run the
Tools
→
Edit/Compare Configuration
utility and
click the
Write All
button or select the checkbox in the
value column and click
Write Checked
button to write the
changes to the meter.
•
Save the configuration file.
Maintenance and Troubleshooting Manual
Troubleshooting
00809-0100-3812
April 2022
Rosemount 3812 Liquid Ultrasonic Flow Meter
19
Содержание Rosemount 3812
Страница 4: ...Contents Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 4 Emerson com Rosemount ...
Страница 12: ...Routine Maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 12 Emerson com Rosemount ...
Страница 36: ...Troubleshooting Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 36 Emerson com Rosemount ...
Страница 126: ...Meter maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 126 Emerson com Rosemount ...
Страница 130: ...Engineering drawings Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 130 Emerson com Rosemount ...