Figure 3-56: Liquid Ultrasonic Flow Meter electronics
A. Terminal end of Transmitter Electronics Enclosure
B. Backplane board location
C. End cap security latch
D. Base enclosure with Acquisition Module
3.9.1
Replace the CPU Module
Procedure
1. Disable electrical power to the meter.
2. Refer to
for the tools required to complete this procedure.
3. Disconnect security seals on the Transmitter Electronics Enclosure, loosen the end
cap security latches using a 3 mm Allen wrench and remove both end caps from the
Transmitter Electronics Enclosure.
4. If replacing the CPU Module (located closest to the terminal end of the enclosure) or
the Optional I/O Module, use a 3mm flat blade screw driver and disconnect the CPU
Module terminal blocks (or the Optional I/O Module) terminal blocks).
Note
The Optional I/O Module is scheduled for future release.
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812
92
Emerson.com/Rosemount