35. Reattach the Transmitter Electronics Enclosure to the Base Enclosure with the four
hex head bolts and lock washers. Tighten bolts with a 7/16 inch (11mm) wrench.
36. Retighten or reattach the conduit to the Transmitter Electronics Enclosure using a
crescent wrench or channel lock pliers.
37. Apply electrical power to the meter.
38. Connect to the meter with MeterLink and run the Zero Flow Calibration Wizard;
select the
Calibrate
→
Zero Calibrate
menu and generate the report. Then, collect
and save a Maintenance log to show the “as left” condition of the meter. Or use
AMS
™
Device Manager or a HART
®
Field Communicator and select
Overview
→
Zero Calibration
→
Start menu
or
Service Tools
→
Maintenance
→
Zero
Calibration
→
Start
.
39. Close your application connection (MeterLink, AMS
™
Device Manager or HART
®
Field Communicator).
40. For band shroud meters only, re-inspect the transducer cables being held in the
body top cable grooves and body circular grooves and re-inspect the cable loop at
the body transducer pocket. Insure that the transducer cables will not be pinched
when the band shrouds are installed.
41. For split shroud meters only, inspect the transducer cables to be sure they will not
be pinched as the lower shroud is assembled. Make sure the transducer cables are
moved out of the way and will not be pinched when the Shroud is reinstalled.
42. Reinstall the shroud:
•
•
•
•
43. Reseal the meter electronics and shroud, if required (see
).
This completes the transducer installation procedure for meters with direct mount
transmitter electronics.
3.7.2
Replace transducers remote mount transmitter
electronics
The meter has integrally mounted transducer cables that require you to disconnect the
cables from the terminal blocks inside the Junction Box and remove the transducer capsule
from the housing. This includes removing the Junction Box assembly from the meter body
and disconnecting the terminal blocks and transducer cable wiring.
Tip
The transducers are installed in matched pairs. Always replace the transducer pair of equal
path lengths (A1 and A2 or B1 and B2). Also, if a transducer pair is moved from one chord
position to another, the transducer pair’s delay time and delta delay time configuration
parameters must be properly re-configured.
Maintenance and Troubleshooting Manual
Meter maintenance
00809-0100-3812
April 2022
Rosemount 3812 Liquid Ultrasonic Flow Meter
77
Содержание Rosemount 3812
Страница 4: ...Contents Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 4 Emerson com Rosemount ...
Страница 12: ...Routine Maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 12 Emerson com Rosemount ...
Страница 36: ...Troubleshooting Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 36 Emerson com Rosemount ...
Страница 126: ...Meter maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 126 Emerson com Rosemount ...
Страница 130: ...Engineering drawings Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 130 Emerson com Rosemount ...