18. Insert the transducer capsule into the transducer housing inside the meter body.
19. Slide the spring, shoulder washer and retaining ring onto the transducer cable.
Insert these components into the transducer housing.
Figure 3-50: Transducer cable assembly
A. Feed-thru cable seal
B. Retaining ring
C. Shoulder washer
D. Spring
E. Transducer capsule
F. Transducer housing
20. Insert the Feed-thru cable seal into the housing groove with the Retaining Ring
pliers (see
).
21. For band shroud meters only, loop the transducer cable at the transducer a couple
of turns. The loop must fit within the body transducer pocket.
a) Place the cable into the circular body groove at the transducer.
b) Use foil tape to hold the cable and loop in position.
c) Continue positioning the cable in the body circular groove and securing with
the tape. At the meter top, position the cable in the body top cable groove
and tape to hold.
d) Ensure that a flat surface will be available for the band shroud to be installed
without squeezing the transducer cable. Repeat for each transducer being
replaced.
Meter maintenance
Maintenance and Troubleshooting Manual
April 2022
00809-0100-3812
84
Emerson.com/Rosemount
Содержание Rosemount 3812
Страница 4: ...Contents Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 4 Emerson com Rosemount ...
Страница 12: ...Routine Maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 12 Emerson com Rosemount ...
Страница 36: ...Troubleshooting Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 36 Emerson com Rosemount ...
Страница 126: ...Meter maintenance Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 126 Emerson com Rosemount ...
Страница 130: ...Engineering drawings Maintenance and Troubleshooting Manual April 2022 00809 0100 3812 130 Emerson com Rosemount ...